105
Calibration
,
continued
Run Cal,
continued
7. Maintenance
3. Select
MENU > CALIBRATION > Run O
2
Cal
to perform O
2
Cal.
Sensor status
1. The sensor status from the latest calibration performed appears as
blue (normal) or red (Slope Error).
Note:
Parameters that are not selected in the
Setup-Parameter
are
marked in grey.
Parameters that are not supported by the inserted cartridge will be
marked without color indicating as inactive status.
2. When pressing a sensor status icon twice, the slope of the
corresponding sensor obtained from the latest calibration appears.