Step 8

Obtain a blood sample using the lancing 

device. Place the device against the pad 

of the finger. The best puncture sites are 

on the middle or ring fingers. Press the 

release button. Remove the device from 

the finger. Wait a few seconds for a blood 

drop to form. A minimum volume of 0.5 microliter is needed to 

fill the confirmation window. (Actual size of 0.5 µL :    )

Applying Blood Sample

Note:

 

The meter may switch off if the blood sample is not applied 

within 2 minutes of the   symbol appearing on the screen. 

If the meter turns off, remove the strip, reinsert it and apply 

blood sample after   symbol appears on the screen.

Step 9

After the   symbol appears on the screen, apply the blood 

sample to the narrow end of the test strip till the meter beeps. 

If the confirmation window is not filled in time because of 

abnormal viscosity (thickness and stickiness) or insufficient 

volume, the Er4 message may appear. 

It is recommended that the application of blood sample to the 

test strip be performed virtually vertical to the sample site as 

shown in the diagram below.

check

Good Sample

Insufficient 

Sample

Step 10

The test result will appear after the 

meter counts down from 5 to 1. The 

result will be automatically stored in 

the meter’s memory.

If the test strip is removed after the 

test result is displayed, the meter 

will automatically switch off after 3 

seconds. Discard used test strips safely 

in disposable containers.

Summary of Contents for TEE2

Page 1: ...i SENS Inc AllRightsReserved PGC8E3060REV003 2015 Medical Technology Promedt Consulting GmbH Altenhofstrasse 80 D 66386 St Ingbert Germany TM Spirit Healthcare Ltd Spirit House Selbury Drive Oadby Lei...

Page 2: ...ntrol Solution Test Results Testing Using the Lancing Device Preparing the Lancing Device Preparing the Meter and Test Strip Marking Post meal Test Results Applying Blood Sample Discarding Used Lancet...

Page 3: ...in just 5 seconds with a small 0 5 L blood sample WelcometotheTEE2 BloodGlucoseMonitoringSystem No part of this document may be reproduced in any form or by any means without the prior written consen...

Page 4: ...code related errors in monitoring by using the no coding function The TEE2 Blood Glucose Meter should be used only with the TEE2 Strip An abnormally high or low red blood cell count hematocrit level...

Page 5: ...ange 1 1 33 3 mmol L Sample size Minimum 0 5 L Test time 5 seconds Sample type Fresh capillary whole blood Calibration Plasma equivalent Assay method Electrochemical Battery life 5 000 tests Power Two...

Page 6: ...aution Do not expose the meter to direct sunlight heat or excessive humidity for an extended period of time Do not let dirt dust blood or water enter into the meter s test strip port Do not drop the m...

Page 7: ...cool and dry place at a temperature of 1 30 C 34 86 F Keep test strips away from direct sunlight or heat and do not freeze Store test strips only in their original vial Close the vial tightly after t...

Page 8: ...to be replaced Post meal test mark appears during post meal testing and when post meal test results are displayed mmol L unit for measuring blood glucose Mute symbol appears only when the sound is set...

Page 9: ...st until the correct year appears Press and hold C button to scroll through the numbers quickly After setting the year press the M button to confirm your selection and progress to setting the month St...

Page 10: ...st strip is inserted in the meter When the blood sample is absorbed into the test strip and the test starts When the test result is displayed When you push the M button or C button to turn on the mete...

Page 11: ...d store at a temperature of 8 30 C 46 86 F Manufacturer i SENS Inc 27 36 Gwangun ro Nowon gu S Discard Date LOT PGA2 Glucose Contro TM You may do a control solution test When you want to practice the...

Page 12: ...rinted on the label of the test strip vial Repeat the control solution test if the test result falls outside of this range Out of range results may occur due to the following factors If results contin...

Page 13: ...ing System or any other medically approved lancing device The lancing device should not be used by more than one individual Ensure the lancing device is not shared among different users Use a soft clo...

Page 14: ...three 3 is recommended Step 6 Insert a test strip with the contact bars facing upwards into the meter s test strip port Push the strip in gently until the meter beeps Be careful not to bend the test s...

Page 15: ...e after symbol appears on the screen Step 9 After the symbol appears on the screen apply the blood sample to the narrow end of the test strip till the meter beeps If the confirmation window is not fil...

Page 16: ...calves can reduce the pain during testing This method of testing with different parts of the body is called Alternative Site Testing While AST may reduce the pain during testing it may not be simple f...

Page 17: ...recognized When insulin has the biggest effect Two 2 hours after an insulin injection AST Precautions Do not ignore the symptoms of hyperglycemia or hypoglycemia When the results of the test do not r...

Page 18: ...the Spirit Healthcare 24 7 Freephone Support Line at 0800 881 5423 or go to www spirit healthcare co uk if such messages are displayed even if you do not have hyperglycaemia or hypoglycaemia Target Bl...

Page 19: ...he bottom of the screen for 2 seconds followed by the average value and the number of the test results saved within the current day ViewingTest Results Stored in the Meter s Memory Step 2 Press the C...

Page 20: ...back through the averages seen previously Press the M button for 3 seconds to turn the meter off C button Pre meal average 1 7 14 30 90days M button C button Pre meal average 1 7 14 30 90days Pre meal...

Page 21: ...off the PP2 alarm press and hold C for 3 seconds The PP2 character the alarm symbol and then the OFF character will appear on the screen Then the screen will change automatically to the memory check...

Page 22: ...leshooting Check whether the test strip is inserted with the contact bars facing up Check if the strip has been inserted completely into the test strip port Check if the appropriate test strip was use...

Page 23: ...of sample 122 Range tested 2 2 32 1 mmol L Within 0 28 mmol L Within 5 Within 10 Within 15 Within 20 47 111 42 89 111 80 109 111 98 111 111 100 11 11 100 Accuracy results for glucose concentration 4...