Note
The function “automatic shut-down” is a protection against battery discharging. When the
device is not active it will be shut down automatically after the time set here. The device
will not be shut down, when data capturing is active or when an external power supply is
applied via USB or DC.
The function is switched off, when the time is set to zero.
sampleRate_ms:
sampling rate in ms
fileType:
possible configuration "dat" or "csv"
trigger:
possible configuration "none" or "highPulse"
autoCleanup:
on
: At least 100 MB of hard disk memory space remain free by
deleting the oldest measuring files. The recording of measuring
data continues.
off
: If free memory space on the internal hard disk is below
40 MB, the recording of measurement data will be automatically
cancelled.
Default setting:
„off“.
Note
Only files with the following format are deleted by the “auto cleanup” function:
paduc_2014-12-31_23.59.59.dat
Only empty directories with the following format are deleted:
2014-12-31
name:
Names can be assigned to the 8 analog and 8 digital input sig-
nals. These names will be visualized for example in the applica-
tion ibaAnalyzer. The name may have up to 63 characters. The
default name for analog
signals is „Voltage[0]“ to „Voltage[7]“,
for the digital signals „Digital[0]“ to „Digital[7]“.
unit:
Unit of the value to be measured
The maximum length of the name is 7 characters
Example: W (power given in Watt)
max:
Maximum value, used for scaling
2
Example: 10 V input voltage = 100 W
min:
Minimum value, used for scaling
Example: -10 V input voltage = -100 W
It is also possible to scale the signals asymmetrically:
Example: 10 V input voltage = 250 A
-10 V inputs voltage = -100 A
6.
Save the configuration file again under "config.ini" in the "config" directory.
2
The scaling factors can be entered according to scientific notation, i.e. 10
3
complies with 1 E3. If invalid
values are entered (letters except E or special characters), scaling will be ignored.