Issue 1.11 

 

 

6.5 

Firmware update 

 

Important Note 

If you are performing a Firmware update, connect the device to an external power sup-
ply! The update could fail if the battery is too week. 

Do not switch off the device during the update process. 

 

6.5.1 

Update via USB 

You can perform a Firmware update by saving the update file onto the internal memory 
of the device and restart. Proceed as follows: 

1. 

Connect the device via the USB cable to the PC. The device will boot up automati-
cally.  

2. 

The device appears on the PC as a USB mass storage. 

3. 

Open the symbol in explorer and copy the update file to the device memory in the 
"update" directory. 

4. 

Close the explorer and remove the device symbol from the system bar. 

5. 

Restart the device by pressing the S11 key longer than 3 s. All LEDs then extinguish 
and after a brief period the device restarts. 

6. 

The update is being performed and all LEDs flash orange. The process can last sev-
eral minutes. 

7. 

The device will restart once the update installation is ended. 

 

6.5.2 

Update via FTP 

1. 

Enter the device IP address or the hostname as server address. 

  Notes, regarding the hostname nomenclature, see chap. 6.4.2 

2. 

Copy the update data into the "update" directory. 

3. 

Restart the device by pressing the S11 key longer than 3 s. All LEDs then extinguish 
and after a brief period the device restarts.  

4. 

The update is being performed and all LEDs flash orange. The process can last sev-
eral minutes. 

5. 

The device will restart once the update installation is ended. 

Summary of Contents for ibaPADU-C-8AI

Page 1: ...ibaPADU C 8AI Mains Independent Data Logger Manual Issue 1 11...

Page 2: ...d trademark of the Microsoft Corporation Other product and company names mentioned in this manual may be labels or registered trademarks of the corre sponding owners Certification The device is certif...

Page 3: ...mory 13 6 Commissioning and configuration 14 6 1 Requirements 14 6 2 Configuration 14 6 2 1 Time synchronization 17 6 2 2 Product information 18 6 3 Start measurement process 19 6 3 1 Start measuremen...

Page 4: ...capable of assessing safety and recognizing possible consequences and risks on the basis of his her specialist train ing knowledge and experience and knowledge of the standard regulations 1 2 Designat...

Page 5: ...ess and the system or machine to be controlled there is a risk of death or severe injury The non observance of this safety information may result in a potential risk of death or severe injury The non...

Page 6: ...onnected to a PC via an USB connection The PC recognizes the devices per plug and play like a mass storage device In addition it is possible to retrieve the data using the network connectivity via FTP...

Page 7: ...lable Strain relief plate Manual 4 Safety instructions Maintaining the operational voltage range Operate the device only with a voltage of 9 30 V DC Too high of an operational voltage will destroy the...

Page 8: ...C 8AI 8 Issue 1 11 5 Description of the device 5 1 Device overview 1 Operational state display LEDs L1 L4 2 S11 key 3 DC input X14 4 Ethernet interface X22 5 USB interface X23 6 Analog digital inputs...

Page 9: ...es orange Lights up red Battery is almost empty Flashes red Error status All LEDs light up red Boot process All LEDs flash orange Update is being loaded All LEDs are off Not in operation no power supp...

Page 10: ...not connected as a mass storage via USB and an IP address has been configured 5 4 3 USB interface X23 A PC can be connected via this USB interface in order to read data At the same time the internal b...

Page 11: ...the analog digital inputs E 1 2 E 10 V A0 3 4 A4 10 V GND 5 6 GND A1 7 8 A5 GND 9 10 GND A2 11 12 A6 GND 13 14 GND A3 15 16 A7 GND 17 18 GND GND 19 20 GND D0 21 22 D4 D1 23 24 D5 D2 25 26 D6 D3 27 28...

Page 12: ...is identical to the digital input design There are 2 ways to con nect the trigger signal Signal level against GND Signal level Trigger 2 2 V active 0 9 V signal level 2 2 V undefined 0 9 V inactive ex...

Page 13: ...d with the following directory structure Directory Contents app ibaAnalyzer application config Configuration file config ini Example of configuration file template_config ini dat Directory of the meas...

Page 14: ...hernet The following parameter can be set sampling rate trigger time zone hours and minutes data format dat or csv DHCP device IP address subnet mask gateway IP address for time synchronization signal...

Page 15: ...5 Possible settings dhcp This setting determines whether the IP address of the device should be assigned by a DHCP server in the network ipAddress IP address of the device This setting will be ignored...

Page 16: ...ted by the auto cleanup function paduc_2014 12 31_23 59 59 dat Only empty directories with the following format are deleted 2014 12 31 name Names can be assigned to the 8 analog and 8 digital input si...

Page 17: ...log directory which indicates the current error in clear text 6 2 1 Time synchronization The time can be synchronized via an external server The timeserver address is stored in the device default time...

Page 18: ...lable measurement will be started without synchronization 6 2 2 Product information The file product log is stored on the device in the log directory It contains important device information such as h...

Page 19: ...e trigger signal must be pending at least as long as the configured sample rate in order to start the data recording The trigger signal connection is described in chapter 5 4 4 Analog digital inputs X...

Page 20: ...ose the explorer and remove the device symbol from the system bar with Securely remove hardware Note The measurement data should be deleted after successfully transmitting to the PC in order to always...

Page 21: ...sists of the product name and the 6 digit serial number of the device You find the serial number on the type label on the device Example serial number 000001 Hostname paduc_000001 The hostname can als...

Page 22: ...ctory 4 Close the explorer and remove the device symbol from the system bar 5 Restart the device by pressing the S11 key longer than 3 s All LEDs then extinguish and after a brief period the device re...

Page 23: ...as residual waste For proper disposal or exchange send the complete device via post using sufficient stamps back to iba The batteries are manufactured with components made of valuable materials which...

Page 24: ...ge Digital inputs Number 8 Version Single ended no galvanic isolation Input signal range 0 V to 30 V Signal level log 0 0 9 V Signal level log 1 2 2 V Sampling rate Linked with analog sampling Communi...

Page 25: ...Operational and ambient conditions Operational temperature range 32 F to 122 F 0 C to 50 C Storage temperature range 4 F to 140 F 20 C to 60 C Transport temperature range 4 F to 140 F 20 C to 60 C Hum...

Page 26: ...onnection plug arrangement of the analog digital inputs PE 1 2 PE A0 3 4 A4 GND 5 6 GND A1 7 8 A5 GND 9 10 GND A2 11 12 A6 GND 13 14 GND A3 15 16 A7 GND 17 18 GND GND 19 20 GND D0 21 22 D4 D1 23 24 D5...

Page 27: ...al number iba S N of the product Contact Headquarters iba AG Koenigswarterstr 44 90762 Fuerth Germany Phone 49 911 97282 0 Fax 49 911 97282 33 Email iba iba ag com Contact Mr Harald Opel Shipping addr...

Reviews: