20
OPERATION
Without timer mode
1. Unit power up
• Connect the AC line cord to a grounded, 3-prong wall outlet.
• Connect the unit to the electrophoresis device using the safety connecting
leads supplied with the electrophoresis device.
• Turn on the power switch located on the rear panel.
• The unit is automatically in STOP mode and the STOP LED
(#11)
will be lit.
The voltage display
(#4)
and current/timer display
(#3)
indicate “000”.
2. Adjusting Output set Values. Constant Voltage
• Press SET/OUT switch
(#1)
to select the SET mode.
At first, the Volts LED
(#5)
blinks.
• Use switches
(#8)
for selection of the voltage value.
• Press V/mA switch
(#2)
for selection of mA parameter. The mA LED
(#6)
blinks.
• Use switches
(#8)
to increase current to the maximum value; 500mA
to avoid constant current mode.
3. Adjusting Output set Values. Constant Current
• Press SET/OUT switch
(#1)
to select the SET mode. Press the V/mA
switch
(#2)
until the mA LED
(#6)
blinks.
• Use switches
(#8)
to select the desired current value.
• Press V/mA switch
(#2)
for selection of Volt parameter. The Volt LED
(#5)
blinks.
• Use switches
(#8)
to increase Volts to the maximum value; 300 Volts
to avoid constant voltage mode.
4. START mode
• Press START, the LED
(#10)
lights and the actual values are displayed
on each display, left for voltage right for current.
• The actual constant mode is indicated by one of the LED
(#5 or #6)
.