5. Changing the settings when unit is working
It is possible to change the settings when the unit is working.
Proceed as follow:
• It is not necessary to press the STOP switch, leave the unit in START mode.
• Press SET/OUT switch
(#1)
to select the SET mode.
• Press V/mA switch
(#2)
to select Volt or mA depending on which parameter
you need to adjust.
• Proceed as described in the above paragraph to change the settings.
• Once the new settings are entered, the unit will display the actual values
after 10 secondes.
Operation with timer mode
and “Gel Saver“ function
1. Unit power up
• Follow the same instructions described on page 21, paragraph 1. 2. 3.
2. Timer selection
• Press ON/OFF switch located in the timer area
(#12)
.
• The LED TIMER
(#13)
lights, showing that timer mode is selected.
• Press clear switch
(#14)
to cancel the previous elapsed time.
• Press SET/OUT switch
(#1)
to select the SET mode.
• Press the time switch
(#15)
until the minutes LED
(#13)
blinks.
• Use switches
(#8)
for selection of the desired separation time.
3. GEL SAVER mode
• This mode is possible when timer mode is selected.
• Press “GEL SAVER” switch
(#16)
.
• The LED “GEL SAVER”
(#17)
lights. The mode is ON until the STOP switch
is pressed.
- When set time has elapsed, the unit turns automatically to constant
voltage and supplies 10% of the voltage set at the beginning of the run,
- The mA/minute display
(#3)
shows the elapsed time,
- The voltage display
(#4)
shows a value, indicating that 10% of the set
voltage is supplied into the electrophoresis tank:
21