Example: 150 Volts was set at the beginning, 15 volts will be supplied when
gel saver function will start to work after timing mode.
• When “GEL SAVER” is working, the LED
(#17 )
blinks. 5 beeps are heard
every minute.
• To stop the “GEL SAVER” mode, press the stop switch
(#9)
.
• It is possible to cancel the “GEL SAVER” option during the run by simply
pressing the switch
(#16)
. The “GEL SAVER” LED
(#17)
turns off.
• It is possible to select the “GEL SAVER” mode during the run by simply
pressing the “GEL SAVER” switch
(#16)
. The LED
(#17)
turns on.
4. Adjusting Output set values. Constant Voltage
• Follow the same instructions described on page 20, paragraph 2.
5. Adjusting Output set values. Constant Current
• Follow the same instructions described on page 20, paragraph 3.
6. START Mode
• Press START, the LED
(#10)
lights and the actual values are displayed
on each display, left for voltage, right for current or minutes.
• To read minutes instead of mA or vice versa, press the TIME switch
(#15)
or VOLTS/mA switch
(#2)
.
• The actual constant mode is indicated by one of the LED
(#5 or #6)
.
• When set time has elapsed, the unit will turn on “GEL SAVER” option
if this option has been selected, or in STOP mode with an audible alarm.
• If this option has been selected, and when set time has elapsed, the unit
turns automatically to constant voltage and supplies 10% of the voltage
set at the beginning of the run,
- The mA/minute display
(#3)
shows the elapsed time,
- The voltage display
(#4)
shows a value, indicating that 10% of the set
voltage is supplied into the electrophoresis tank.
7. Changing the settings
• Follow the same instructions described on previous page 21, paragraph 5.
22