Manual de Operação - SONOPEEL
1
a
edição (10/2009)
34
Field destined to the choice of the parameter
APPLICATION TIME
(TIMER) which can range from 1 to 30 minutes.
Field destined to the choice of the parameter
CURRENT TYPE
which
allows the selection of electrical stimulation by microcurrent (MENS) or
microgalvanic (GMES).
Field destined to the choice of the parameter
POLARITY
which allows
the selection of microgalvanic current polarity (GMES):
(P+)
positive
polarity and
(P-)
negative polarity.
Field destined to the choice of the parameter
FREQUENCY
(Hz) which
refers to the pulse repetition frequency. Enabled only for microcurrent
(MENS) and ranges from 1 Hz to 500 Hz.
Field destined to the choice of the parameter
CURRENT INTENSITY.
It indicates the current intensity in µA (micro ampères) for the current
MENS and GMES. Ranging from 0 to 990 µA (“steps” of 10 µA).
1st step:
Remove the equipment from the transport box and from the plastic bag
protection. Connect the detachable power supply cable (12) and connect the equipment
into the local electrical network socket.
2nd step:
To perform the
ULTRASONIC PEELING
or
SONOPHORESIS
or
COMBINED THERAPY
or
ISOLATED THERAPY
, connect the spatula
transducer into the connection (11) located in the front part of
SONOPEEL.
The blue
channel is for the connection of the ultrasound peeling spatula transducer and the
fixation connector of the spatula handle has screws which must be fixed into the output
connector. Always tighten the fixation screws of this connector.
3rd step:
When choosing the
COMBINED THERAPY
or
ISOLATED THERAPY
also place the connection cable for electrostimulation into the output connector
(CURRENT) located in the front part of the equipment.