Manufacturer’s guidelines and declaration – Electromagnetic immunity
SONOPEEL is destined to be used in the electromagnetic environment specified
below. The user of the equipment should ensure that it will be used in such an
environment.
Immunity test
Test level
IEC 60601
Conformity level
Electromagnetic
environment- guidelines
RF Conducted
IEC 61000-4-6
RF Radiated
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
10 V/m
80 MHz to
2,5 GHz
3 V
10 V/m
Communication equipment of
RF portable or mobile should
not be used near any part of
SONOPEEL,
including
cables,
with
separation
distance smaller than the
recommended,
calculated
from the equation applicable
to the transmitter frequency.
Recommended
separation
distance
d = 1,2
P
d = 0,35
P
80 MHz to 800
MHz
d = 0,7
P
800 MHz to 2,5
GHz
Where P is the maximum
nominal
output
of
the
transmitter in watts (W),
according to the transmitter
manufacturer, and d is the
recommended
separation
distance in meters (m).
It is recommended that the
field intensity established by
the
RF
transmitter,
as
determined
through
an
electromagnetic inspection on
the local,
a
be smaller than the
conformity level in each
frequency band
b
.
Interference
may
occur