7.4 FAQ
Current questions and answers can be found on our website.
Independent triggering of keys
Appliances that work with strong electromagnetic fields, such as mobile
phones, electrotomes, etc. can trigger malfunctions in direct proximity
to the keyboard. Only operate such appliances at a large distance from
the keyboard.
Individual keys do not work
When changing the batteries (at the moment when the battery compart-
ment was closed) keys were pressed by mistake.
Open the battery cover and then ensure that you do not touch any key
for several seconds when you close the battery cover. The keyboard
needs this period of time to calibrate itself correctly.
Individual keys no longer work
correctly
Check the battery charge status and change the batteries as necessary.
Open and close the battery compartment (reset).