Wiring the Gateway to the Chillgard LC
Connect the “RS-485 To Sensor Module” terminals
−
, +, and G on the Chillgard LC to the gateway’s terminal
block at terminals TB:2 (B), TB:1 (A), and TB:6 (GND) as shown in Figure 2. This terminal block on the Chillgard
LC will have wires for both the gateway and any sensors connected to it. Additionally, a jumper wire must be
placed on the gateway’s terminal block between terminals TB:1 (A) and TB:3 (Y). Another jumper wire must be
placed between terminals TB:2 (B) and TB:4 (Z). as shown in Figure 2.
Figure 2: Gateway and Chillgard LC Connections