9.1 Reset for overtemperature
The equipment has a separate safety device to prevent that the heating element tem
-
perature rises uncontrolled, for example, when it runs out of water.
In that case, the mechanism disconnects the heating element power and, once cooled,
it must be reset in order to continue using the computer.
Press the reset button on the back side of the computer.
9.2 Setting the «Offset» temperature reading
The offset (A) and the ºC/ºF unit (B) can be adjusted.
A. Adjust the probe reading correction (offset)
Press keys (2) and (3) continuously and at the same time (about 5 seconds).
The
OFFs
message will appear.
Use keys (2) / (3) to set the actual temperature reading value. Press key (1)
to confirm.
B. Presentation unit: ºC / ºF
Press keys (2) and (3) continuously and at the same time (about 5 seconds).
The
OFFs
message will appear.
Press the key (4). The message
Unit
will appear. Using keys (2) / (3), select the
unit: ºC / ºF. Confirm with key (1).
10. WARRANTY
This product is guaranteed for one year. It does not cover damages caused by an incorrect
use or causes beyond the control of ICC.
Any manipulation of the equipment by personnel not authorized by ICC automatically
cancels the guarantee.
Description
Cause
Resolution
It does not heat
The safety thermostat has skipped
Press the thermostat reset. See 9.1
Heating element is melted
Contact the technical service to replace
Temperature reading does
not match the external ther
-
mometer
Temperature imbalance
See 9.2 on how to act for adjustment
9. ERROR MESSAGES & TROUBLESHOOTING
P.FAIL
Power failure. The equipment continues with the process.
Err
will appear in green on the top line of the display.
(Note: if you turn off the device without pressing STOP, this error will appear when
the device is restarted)
TSEG
The safety thermostat is open.
SENS
Open temperature probe.
The control does not detect the temperature probe.
HEAT
The resistance does not work.
The equipment detects that the temperature does not increase.