MANUAL DE INSTRUCCIONES
CODIGO 80300 REV C
09/2020 Pag.: 9
J
.P. SELECTA s.a.u.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.grupo-selecta.com
Electric risk
Risk of high
temperature contact
7. MAINTENANCE & CLEANING
MAINTENANCE:
Before operating inside, disconnect from the mains.
The manipulation of the equipment internal electronic circuits by unauthorized
personnel can cause irreparable damage. Take it to one of ICC’s authorized
technical services.
CLEANING:
To clean the accessible parts of the equipment, use a cleaner or a similar product.
Do not use aggressive solvents. To remove lime, you can use vinegar a bit lowered.
8. TECHNICAL SPECIFICATIONS
Reference code:
80013
80015
Dimensions: height/width/depth (mm)
280/180/190
Supply voltage:
230 VAC 50 - 60 Hz
115 VAC 50 - 60Hz
Electrical power (W):
2060
1060
Weight: (Kg)
4
External fuse:
Replaceable (Type: 10x5mm)
Temperature range: (No refrigeration)
Ambient Tª + 5ºC ...
150ºC
Display resolution:
0.1ºC
6.4 Circulation pump directions
The water circulation pump could operate in two configuration modes:
-
INTERNAL circulation
:
Water will be pumped from horizontal pump output (See fig 6.4.1).
Use this configuration to made water circulation into the bath vessel in order to homo
-
genize the temperature from sample load.
-
EXTERNAL circulation
:
Water will be pumped from vertical pump output (See fig 6.4.2).
Use this configuration to made water circulation out of the bath vessel in order to tem
-
perate an external equipment.
Fig 6.4.1
INTERNAL circulation
.
Fig 6.4.2
EXTERNAL circulation
.