Recommended separation distances between portable and mobile RF communications
equipment and the EQUIPMENT or SYSTEM- for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and
the model IceSense3 Cryotherapy product
The model IceSense3 Cryotherapy product is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the model IceSense3
Cryotherapy product can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the model IceSense3
Cryotherapy product as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.