IceSense3™
2.4.6
Grounding
Warning
To avoid risk of electric shock, this equipment must only be connected to a supply main with
protective earth.
2.4.7
Sterility
Warning
The cryoprobes and the temperature sensors are single use and are supplied in single use
packaging. Never reuse a single-use cryoprobe, temperature sensor, or sterile sleeve.
Reprocessing single use device (like the cryoprobe, temperature sensor) affects the mechanical,
performance, and microbiological properties of the product.
Warning
After finishing the surgical procedure on a patient, remove and discard the single-use cryoprobe,
temperature sensor, and sterile sleeve.
Warning
For each new patient, ensure that the previously used single-use cryoprobe, temperature sensor,
and sterile sleeve have been removed and discarded. Any used cryoprobe and temperature
sensor should be related as used sharp biohazard waste.
2.4.8
Mechanical handling of flexible hose and cryohandle
Warning
Never use excessive force to insert or remove a cryoprobe from the cryohandle.
If moderate force is not sufficient, contact IceCure Medical for advice.
See 8.2
See 3.3.5
See 5.2.7