User Manual
22
IceSense3™
Caution
Do not pull the system by the cryohandle or flexible hose.
The flexible hose shouldn’t be bent or flexed during procedure.
2.5
Emergencies and errors
2.5.1
Emergency Stop button
Caution
Only push the Emergency Stop button when there is NO other choice. Whenever possible, use
Warm to release the probe from the tissue and use the standard shutdown procedure. Excessive
use of the Emergency Stop button may damage the system.
Warning
If a liquid nitrogen leak is detected during the cryotherapy procedure, PUSH the Emergency Stop
button immediately.
2.5.2
Emergencies causing procedure halt
Warning
When a procedure halts due to an error, switch off the system. Call IceCure Medical and describe
the error shown on the screen as precisely as possible.
Do NOT attempt to reuse the system before contacting IceCure Medical.
Warning
When an error message is displayed and a procedure is aborted, remove the probe ONLY after
you are allowed to do so by the system. Failure to do so will increase the risk of liquid nitrogen
related accident.
See 3.3.3
See 5.5.2
See 5.5.
See 6.3.3