34
IceSense3™
Figure 19: The Emergency Stop button
Caution
Only push the Emergency Stop button when there is NO other choice. Whenever possible,
use Warm to release the probe from the tissue and use the standard shutdown
procedure. Excessive use of the Emergency Stop button may damage the system
.
Warning
If a liquid nitrogen leak is detected during the cryotherapy procedure, PUSH the Emergency Stop
button immediately.
3.3.3
Flexible hose
The flexible hose joins the cryohandle to the siphon inside the chassis. This permits nitrogen to flow
from the main chassis to the probe.
Caution
Do not pull the system by the cryohandle or flexible hose.
The flexible hose shouldn’t be bent or flexed during procedure.
3.3.4
Touch screen
The touch screen is located on the top of the main chassis and allows for operating and monitoring
of the system. It is designed for users and technicians.
Do not connect any signal input/output port to the touch panel PC except medical certified
equipment provided by IceCure Medical.
A sample touch screen display is represented by the figure below. For further information about
computer interface and operating the system, see Chapter 6.