When pressing OK, the system will move to Manual mode and be in a Thaw cycle.
Pay attention that in Manual mode, you need to press the icon on the screen for each freeze or
thaw cycle.
Replace Dewar as in paragraph 5.3.8.
If you press cancel, you will end the procedure (see figure below).
Figure 67: procedure stopped due to Low Nitrogen.
5.4
Post-operational stages
5.4.1
Removing the temperature sensor from tissue
At the end of the cryotherapy procedure the temperature sensor should be removed from the tissue
by carefully pulling on it. The extraction should be done only after a passive thaw of a few minutes
since the temperature sensor can be surrounded by ice. Removing it too quickly before thawing can
damage the target tissue.