At the end of the last treatment of the day, Switch OFF the mechanical ON/OFF button and unplug
the electrical cable. Clean the system following instructions in section 8.1 and move the system to its
storage location.
5.5
System failures
5.5.1
IceSense3™ cryoablation system failure
When the IceSense3™
cryoablation system detects an error, the following will occur:
A
failure message
will appear.
The procedure will be aborted.
When the above occurs, take action as follows:
1.
Write down the error message and number, follow the system instructions
.
Figure 73: System Fail example screen
2.
If you are in the middle of a cryotherapy procedure,
wait for passive thaw
and then carefully
remove the cryoprobe
from the tissue.
3.
Detach the cryoprobe
from the cryohandle ONLY after Dewar the system instructs you that it
is safe to do so.
4.
If system shut down is required,
shut off the IceSense3™ cryoablation system
by pressing the
mechanical ON/OFF button on the main chassis and then remove the plug from the power
supply.
5.5.2
IceSense3™ cryoablation system touch screen failure
If the touch screen shuts down, you will need to shut off the system with the mechanical ON/OFF
button and remove the plug from the power supply. This will end the cryotherapy procedure.
Headline
User
Instructions
Error Number