IceCure Medical Ltd.
DSR3000000 rev.L
Confidential
User Manual
94
IceSense3™
Problem
Probable Cause
Action
sensor.
If temperature reading is still
unreasonable, continue the
procedure without a
temperature sensor.
If problem persists, contact
IceCure Medical service.
Cryoprobe cannot
easily be removed from
target tissue.
Warm process malfunction
Do not use force in attempting to
remove the cryoprobe from the
tissue.
Wait a few minutes and try
removing it gently again.
If problem persists, contact
IceCure Medical service.
Cryoprobe is difficult to
detach from
cryohandle.
Parts are frozen together
Try again after five minutes. If it
still cannot be detached, contact
IceCure Medical service.
Nitrogen leak when
trying to remove the
cryoprobe
Excess pressure on the pipes,
main valve open, procedure not
completed, or Dewar relief valve
malfunction.
When initial leak is detected,
screw the cryoprobe back in
place.
Make sure that the procedure
has ended, wait 3 minutes and
try removing the cryoprobe
again. If nitrogen is still leaking
contact IceCure Medical service.
Computer failure due
to hardware fault
Unknown
Mechanical shut down of the
system using the On/Off switch
at the back of the system or
emergency stop.
Then turn on the system (after
releasing the emergency stop
button, if pressed) and wait for
computer re-boot.
Screen is stuck/
Controller stops
working/ Controller is
stuck
Unknown
Mechanical shut down of the
system using the On/Off switch
at the back of the system or
emergency stop.
Then turn on the system (after
releasing the emergency stop
button, if pressed) and wait for
computer re-boot.
Table 1: Troubleshooting guide