Moving the Scanner
Chapter 1 Moving the Scanner
Safety
Warning: The
360
scanner weighs approximately
100 kg (220 lb). A minimum of two
persons are needed for lifting or
moving the unit.
Warning: Do not tilt the scanner more than 10° from
the vertical.
Warning: When moving the scanner more than a
short distance, a suitable trolley must be
used.
Moving Procedure
Once the
360
scanner has been installed you
should rarely need to move it. If you wish to move it
a short distance (for cleaning or accessing the
electrical connections for example), use the
procedure which follows. Keep the scanner on a firm
surface at all times.
Caution: Before moving the scanner its base cover
must be removed to protect it from
damage when the scanner is tilted.
1. Remove the base cover from the scanner to
protect it from damage. To do this:
a. Open the scanner lower door.
b. Using a Phillips screwdriver release the
four, quarter turn fasteners securing the
base cover:
3250360-2
15