D700 | D720 Series Operating Manual
4
Technical Specifications
Measuring Range
pH: -2 – 16pH
|
ORP: -2000 – +2000mV
Unit
pH
|
mV
Resolution
pH: 0.001pH
|
ORP: 1mV
Basic Error
pH: 0.001pH
|
ORP: 1mV
Temperature
14 – 302°F
|
-10 – 150°C
(Sensor Selection Dependant)
Temperature Resolution
0.18°F
|
0.1°C
Temperature Accuracy
±0.54°F
|
±0.3°C
Temp. Compensation
32 – 302°F
|
-10 – 150°C
Temp. Compensation
Manual or Automatic
Stability
pH: 0.01pH/24h
|
ORP: 1mV/24h
Output
Two 4–20mA
|
20–4mA
|
0–20mA
Communication
RS485 MODBUS RTU
Relay Control Contact
5A 240VAC
|
5A – 28VDC
Power Supply
85 – 265VAC
|
9 – 36VDC
|
Power Consumption 3W
Working Temperature
14 – 140°F
|
-10 – 60°C
Relative Humidity
≤ 90%
Waterproof Rating
IP65
Weight
1.5Kg
Dimensions
235 × 185 × 120mm
Installation methods
Wall
Other functions
Data record & Curve display
Dimension