P4
6. After successfully reading the first iWSN-200x register value,
press and hold the Wake button for more than 1 second to
reset to zero.
7. Read the iWSN-200x register and confirm that the value is
close to zero.
PLC
address
(Base1)
Protocol
address
(Base0)
Data
Data type
Description
40257
30257
256
(0x
01
00)
4
Uint16
Mean value: X
40258
30258
257
(0x
01
01)
0
Uint16
Mean value: Y
40259
30259
258
(0x
01
02)
1
Uint16
Mean value: Z
8. Turn off the zero-initialization setting. (set the F1/F2 pin to OFF,
then turn off the transmitter power)
9. Complete the zero setting, please turn on the power again for
vibration measurement.
10. If the measurement position is adjusted, please re-execute
the zero-initialization procedure to ensure the measurement
accuracy.
2
PWR
ON
:
Power on
Item
Name
Instructions
1
Wake
(zero-initialization
setting)
Item
Name
Instructions
1
F1
(Turn off the
zero-initialization setting)
□:
ON
■:
OFF
F2
(Turn off the
zero-initialization setting)
□:
ON
■:
OFF
2
PWR
OFF
:
Power off