Required Equipment
Required equipment is listed in Table B. If the recommended models are not available, equipment with
equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment
Required Characteristics
Recommended Model
Calibrator
AC Voltage:
Range:
0 ~ 1000V AC
Accuracy:
+/- 0.07%
Frequency Range:
40 ~ 100kHz
Accuracy:
+/- 0.01%
DC Voltage
Range:
0 ~ 1000V DC
Accuracy:
+/- 0.006%
AC Ampere
Range:
0 ~ 10A AC
AC mA Accuracy:
+/- 0.15%
AC A Accuracy:
+/- 0.21%
Frequency Range:
40 ~ 30kHz
Accuracy:
+/- 0.01%
DC Ampere
Range:
0 ~ 10A DC
DC mA Accuracy:
+/- 0.0175%
DC A Accuracy:
+/- 0.08%
Frequency Source
Range:
0.1Hz ~ 4MHz
Accuracy:
+/- 0.003%
Resistance
Range:
0Ω ~ 40MΩ
Accuracy:
+/- 0.02%
Capacitance
Range:
1nF ~ 10mF
Accuracy:
+/- 0.5%
Temperature
Range:
0 ~ 1200 °C
Accuracy:
+/- 0.2%
Fluke 5500 or Wavetek
9100 Calibrator or
equipment