1. OVERVIEW

Ideal Vacuum’s Habanero™ is a benchtop sized reference absorbtion cell heating furnace for 

spectroscopy, and can be used with all of our sealed wavelength reference cells. The Habanero 

furnace provides a stable temperature and convenient package for many reference cell applications. 

It provides an even, elevated temperature to maintain a constant absorption spectrum of tellurium 

for laser locking and spectrum calibration. For applications where greater resolution is required, 

it can be used in a saturated absorption system for doppler-free absorption spectroscopy. As 

configured, the Habanero is ready for use in transmission spectroscopy.

With a maximum operating temperature of 900°C, the Habanero furnace is especially useful 

for high temperature references such as tellurium and the excited states of iodine. With its wide 

operating range from maximum to ambient, the Habanero can be used with all of Ideal Vacuum’s 

sealed wavelength reference cells. Note that care must be taken not to exceed the damage 

threshold temperature of alkali and organic reference cells.
Raising the temperature of a reference material that is solid under normal conditions increases the 

number density of atoms or molecules in the gas phase. This increases the optical density of the 

reference, increasing signal-to-noise ratios and the number of lines that can be confidently used 

to calibrate a spectrum. Raising the temperature also changes the distribution of populated states, 

allowing access to calibration-quality absorption lines beginning from higher rotational, vibrational, 

and even electronic states in wavelength ranges inaccessible at normal room temperature. Some 

reference materials have absorption lines or peaks that move as a function of temperature. For 

these materials, the Habanero furnace can provide the precise long-term temperature control 

necessary to maintain laser calibration for the full duration of an experiment.

1.1 INTRODUCTION

Summary of Contents for Habanero P1012726

Page 1: ...HRCF 08122022 V 1 1 USER MANUAL Habanero Spectroscopy Reference Cell Furnace P1012726 ...

Page 2: ...ARTICULAR PURPOSE IDEAL VACUUM SHALL NOT BE LIABLE UNDER ANY CIRCUMSTANCES FOR INDIRECT SPECIAL CONSEQUENTIAL OR INCIDENTAL DAMAGES ARISING OUT OF THE USE OF THE PRODUCT THE TOTAL LIABILITY OF IDEAL VACUUM SHALL NOT EXCEED THE PURCHASE PRICE OF THE PRODUCT UNDER ANY CIRCUMSTANCES Customer Service and Support If you have any questions concerning the installation or operation of this equipment or if...

Page 3: ...mation 5 Warning Symbols and Definitions 5 1 Overview 6 1 1 Introduction 6 1 2 Physical Specifications 7 1 3 Available Wavelength Reference Cells 8 2 Basic Installation 9 2 1 Power and Control 9 2 2 Mounting 9 3 Reference Cell Installation 11 4 Window Replacement 14 4 1 Available Replacement Windows 14 4 2 Window Replacement Procedure 15 5 Furnace Control 16 5 1 With IVP PID Heater Control 16 5 2 ...

Page 4: ...sp and pull on window retaining ring 11 Figure 6 Pull out from end cap 12 Figure 7 Slide off ceramic holder 12 Figure 8 Seat the reference cell into the ceramic holder 12 Figure 9 Slide the ceramic holder back onto the mounting rods 13 Figure 10 Seat the ceramic holder onto the reference cell 13 Figure 11 Insert the cell holder assembly into the furnace 13 Figure 12 Remove screws in window retaini...

Page 5: ... It is used to alert you to potential personal injury hazards Obey all safety messages that follow this symbol to avoid possible injury or death Internationally recognized safety symbols may be used with safety warnings to specify the type of hazard or a safety protocol to follow For example Indicates safety glasses are required Indicates an electric shock hazard Important Safety Information Thank...

Page 6: ...With its wide operating range from maximum to ambient the Habanero can be used with all of Ideal Vacuum s sealed wavelength reference cells Note that care must be taken not to exceed the damage threshold temperature of alkali and organic reference cells Raising the temperature of a reference material that is solid under normal conditions increases the number density of atoms or molecules in the ga...

Page 7: ...lder 139 7 mm 5 5 in Maximum Mounting Rod Sample Diameter 41 3 mm 1 625 in Maximum Cell Length on Mounting Rods 155 mm 6 1 in Heater Sheath Material 304 Stainless Steel Case Material 10 ga 5052 H32 Aluminum Maximum Case Temperature 130 C 266 F Cooling Passive Convection Power Input 100 120 VAC 50 60 Hz Maximum Current 3 6 A Heating Power 425 W Power Cable 118 cm 3 9 ft long with NEMA 15 5 plug The...

Page 8: ...75 8300 nm 10 mm x 75 mm P1012478 Ethane C2 H6 Ambient 3050 3650 nm 10 mm x 75 mm P1012479 Propane C2 H8 Ambient 2950 4000 nm 10 mm x 75 mm P1012480 Ethylene C2 H4 Ambient 3075 3650 nm 10 mm x 50 mm P1012481 Isotopic Acetylene 12 C2 H2 Ambient 2985 3125 nm 10 mm x 50 mm P1012482 Isotopic Acetylene 13 C2 H2 Ambient 2985 3125 nm 10 mm x 50 mm P1012483 Isotopic Carbon Monoxide 12 C16 O Ambient 2300 2...

Page 9: ...or a power strip will result in overheating and permanent damage to the furnace For temperature monitoring the Habanero is equipped with a type K thermocouple with a miniature male straight blade connector 2 Basic Installation 2 2 Mounting For adequate ventilation at least 3 of clearance on each side and 12 above the furnace is required It must be placed on or mounted to a non combustible working ...

Page 10: ...unting to an optical table with standard 1 x 1 breadboard tapped holes we recommend our optical table furnace mounting kit P1012736 shown below which includes everything needed for a vibration free secure installation Other optical table accessories are available including optical posts and holders in varied lengths optic mounts and breadboards Figure 2 Optical post legs for benchtop placement P10...

Page 11: ...rence cell is done when the heater is cold at ambient temperature and requires a 5 64 hex wrench Wear gloves when performing this procedure 1 On the side of the furnace opposite the power cable unscrew the four 6 32 x 3 8 button head screws on the end cap LA 11 405 just inside the furnace case 2 Grasp the window retaining ring and carefully pull the cell holder assembly out of the furnace Figure 4...

Page 12: ...older assembly out 4 Once the cell holder assembly is out of the furnace slide off the first ceramic holder 5 Notice that the end cap optic hole is off center Slide the reference cell in between the rods cell stem pointing towards the larger side of the end cap Seat the cell into the holder Figure 6 Pull out from end cap Figure 7 Slide off ceramic holder Figure 8 Seat the reference cell into the c...

Page 13: ...7 Carefully slide the ceramic holder over and seat it on the reference cell 8 Carefully slide the cell holder assembly back into the furnace cell stem up 9 Secure the assembly to the furnace with the four 6 32 x 3 8 screws Use a 5 64 hex wrench Figure 9 Slide the ceramic holder back onto the mounting rods Figure 10 Seat the ceramic holder onto the reference cell Figure 11 Insert the cell holder as...

Page 14: ...he furnace uses 1 inch diameter windows up to a maximum thickness of 8 mm The windows may be removed entirely which results in reduced heating performance and temperature uniformity Below Table 4 lists our available replacement windows All are 1 diameter 25 4 mm 5 mm 197 in thick 0 5 wedge and have high surface flatness to prevent beam distortion Part Material Maximum Temperature Transmission Rang...

Page 15: ...hen performing this procedure 1 Unscrew the three 6 32 x 3 8 button head screws which secure the window retaining ring LA 11 406 Be careful not to drop the screws or retaining ring into the furnace case 2 Carefully remove the window Do not let it fall into the furnace case 3 Install the replacement window in reverse order 4 2 Window Replacement Procedure Figure 12 Remove screws in window retaining...

Page 16: ...ay will read no Cont 5 Press and hold down the SET button for 5 seconds 6 The display should show CnPt the correct setting for a Type K thermocouple If it does then press the SET button 7 If the controller shows a CnPt value other than that shown in Figure 15 press the UP or DOWN ARROW until it does Then press the SET button to save the setting 8 Once the thermocouple setting is verified turn off ...

Page 17: ...ntroller is factory programmed to turn the furnace on immediately upon power up 15 To turn off furnace heating and enter a new furnace target temperature press the ROTATE button The display will show r S and RUN Figure 18 next page 16 Press the UP ARROW The display changes to STOP Figure 19 next page 17 Press the SET Button to save the selection and go back to the temperature display 18 Similarly ...

Page 18: ...periment turn the controller s power switch off The SV value is retained and as soon as the power switch is turned back on the furnace will immediately begin heating if the r S value is set to RUN Figure 20 Set the furnace target temperature Figure 18 Furnace runs on powerup 5 2 With Variable Transformer If you are controlling the furnace with a variable transformer Variac make sure to use a readi...

Page 19: ...Ideal Vacuum Products LLC 5910 Midway Park Blvd NE Albuquerque NM 87109 Phone 505 872 0037 Fax 505 872 9001 Web idealvac com ...

Reviews: