SE2L User’s Manual (SE9Z-B1865)
113
Figure 7-46
Muting zone configuration example 2
Muting zone configuration example as square is drawn from point 1 to point 2
7.15.4
Reference region configuration
Follow the steps below to configure reference region.
Ensure the SE2L is firmly mounted at its location.
Ensure the connection between the SE2L and PC.
Click the Function tab and enable the Reference Monitor function and enter the required tolerance
distance (default 100mm).
Click the Area tab.
Main screen displays the configuration for reference region (Figure 7-47). First area on the preview
indicated by “REF” represents the reference area.
Configure the
reference region by using the drawing tools. Move the cursor on the drawing panel
and drag and drop to configure the reference. Monitoring region is displayed in green
Protection zone
Muting zone
1
2