Attach your sensor cable to the ToFscan device to get a 

green connection symbol

Prepare and clean the skin for electrodes placement

Check the stimulation current (50mA by default)

2

3

S T A R T - U P

4

Position the splint to the patient

Ensure thumb is sitting flush on the thumb pad and 
the index finger is positionned  in the loop.

5

Ensure the skin is clean and healthy.

Start to measure

7

• 

Press and navigate wheel to the desired mode.

• 

Press and hold for at least 2 seconds the wheel 

button to start the electrical stimulation.

Connection of the sensor 

(green = connected)

Electrode impedance level 

(green = optimum)

Battery charge level

Press the wheel dial to turn the device on

1

Check displayed information

6

Cleaning of the ToFscan

8

Ensure there is battery available

Place  ECG electrodes on the ulnar nerve

• 

Space between electrodes: 2 to 5 cm (0.8in-2in)

• 

Red clamp = proximal electrode.

• 

Black clamp = distal electrode.

2-5 cm

Only a good impedance 

 allows the 

ToFscan to be used in good conditions.

Between each patient, desinfect the device and accessories without 
immersing or spaying them.

Careful not to pull down on the sensor wires while cleaning.

9

Turn off the device

To turn off the ToFscan, navigate wheel to the power symbol and 
hold for 2 seconds