SECTION TWO
–
INSTALLATION AND START-UP
IDRONAUT
–
Brugherio (MB)
OCEAN SEVEN 304
Plus
CTD
12-2015
15
2.
INSTALLATION AND START-UP
Unpacking, installation and start-up procedures are described in this section.
2.1.
SHIPPING LIST
Description
Quantity
- OCEAN SEVEN 304
Plus
1
- Dummy connector with looking sleeve
1
- Laboratory RS232C communication cable
1x 3 m
- 2x 1.5V Alkaline battery holder
1x
- 1x 3.6V Lithium battery holder
1x
- Sensor maintenance kit, including:
Conductivity sensor cleaning solution
1x 25 ml
- CD-ROM programs and documentation
1
2.2.
INSTALLATION PROCEDURE
Unpack and inspect the shipped parts; check the materials with respect to the above shipping list and
be sure that no damages have occurred during the transport. Open the CTD and install the internal
batteries (see dedicated section). Remove the underwater dummy cap from the top cover of the CTD
and connect it with the connector of the laboratory cable. Connect the other side of the cable to a
personal computer USB interface.
2.2.1.
Optional USB-cable software driver installation
The first time the USB laboratory communication cable is connected to a personal computer, the
operator is requested by the Windows operating system to install the USB driver. The driver can be
found on the ITERM or REDAS-5 CD-ROM included in the CTD accessories. The USB driver
installation allows the personal computer and the Windows operating system to communicate with
the Ocean Seven 304
Plus
through a simulated “communication port”. The simulated co
mmunication
port is used by ITERM and other IDRONAUT programs to communicate with the CTD.
2.3.
HOW TO OPERATE THE CTD
The CTD can be operated through the IDRONAUT Terminal Emulation software per Windows (for
brevity’s sake, hereafter referred to as “ITERM”). The “ITERM” software is described in a dedicated
append
ix of this Operator’s Manual. The communication parameters of the softw
are must be set to
meet those of the CTD (see the configuration sheet). The default communication values are:
38400
baud, 8 data bits, 1 stop bit, no parity.
After the installation, t
he “ITERM” programme
must be configured to meet the default
communication set-up. Afterwards, any modification done to the OCEAN SEVEN
304Plus
communication speed must be reflected on the ITERM programme too. ITERM initialization file keeps
trace of the modification for the successive start-up of the ITERM programme.
Run the
ITERM
programme, set it up, and then switch on the CTD by positioning the arm of the
rotating switch on the red dot. The OS
304Plus
now starts to wake up.
Note
At the start-up, the CTD shows you the command prompt, waiting for the oper
ator’s instructions, unless the
CTD has been previously configured to perform self-recording data acquisition cycles. In this latter case, a
dedicated section of the manual describes how to revert the CTD to normal operations.
2.4.
START-UP PROCEDURE
2.4.1.
CTD Start-up - Verbose operating mode
Once the CTD is switched on and the verbose operating mode is operative, the following message
appears:
Summary of Contents for Ocean Seven 304Plus