HTB-100-HM170 Medical Box PC
Page 33
Figure 3-25: Power Button and Power LED
3.14 Clear CMOS
If the HTB-100-HM170 fails to boot due to improper BIOS settings, the clear CMOS button
clears the CMOS data and resets the system BIOS information. To do this, remove the
system top cover first (see
Section 3.3
). Locate the clear CMOS button and push the
button for three seconds, then restart the system. The clear CMOS button location is
shown in
Figure 3-26: Clear CMOS Button
If the “CMOS Settings Wrong” message is displayed during the boot up process, the fault
may be corrected by pressing the F1 to enter the CMOS Setup menu. Do one of the
following:
Enter the correct CMOS setting