20
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 sec
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 sec
operation during
power
mains interruptions,
it is recommended
that the model
POC-W22A-H81 be
powered from an
uninterruptible
power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The model POC-W22A-H81 is intended for use in the electromagnetic environment
specified below. The customer or the user of the model POC-W22A-H81 should
assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Conducted RF
IEC
61000-4-6
Radiated RF
IEC
61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
Vrms
V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the model
POC-W22A-H81, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
d = 1,2
P
d = 1,2
P
80 MHz to 800
MHz
d = 2,3
P
800 MHz to 2,5