Page 22 of 110
Programming Tester Utility Functions
[UTILITY]
Right
1 Prim Param Program
Ls
Connect OPEN standard
Press START to continue
Tester functions are accessed through the UTILITY mode.
Press [UTILITY] at any time to
exit programming mode.
Perform Calibration/Zeroing
Press the
UP
or DOWN
arrow to select
Open/Short Calibration
type
UP
Quick Short Cal, Quick Open Cal,
Quick Short/Open Cal, Short Cal,
Open Cal, Short/Open Full Cal
NOTE: "Quick Cal is not an option if test is a sequence test (S1-S9).
NOTE: "Open Cal" and "Quick Open Cal" are not an option if test is a DC test (DCR, DCV, DCI)
Press [STOP] at any time to
exit Cal routine
Connect Test Leads in OPEN configuration
Press the
[START]
to initiate
open calibration
measurement
C: 1 9: F1 I5 V2
Open correction
[START]
Open Correction factor
Cal with 1M Cable
Cal Due: 1/10/2001
Press the
RIGHT
arrow to skip Open/Short and go to Keypad Lockout with indicated cable length
Cal with 1M Cable
Cal Due: 1/10/2001
Indicates due date of next
annual calibration
Right
To connect Test Leads refer to "Connection to Device Under Test" paragraph that follows.
Kelvin Leads Placed Apart
OPEN
to 1930