Operation/Programming
Page 65 of 110
Program the 1910 for
OPEN
as follows:
Press
START
to initiate
NOTE:
During the open/short process these numbers should be changing.
A flashing fail light during the open/short process indicates an error has occurred.
Remove Open, connect short
Black
Red
Figure 2-5
Short Circuit Configuration (Kelvin Clip Lead Set or Test Fixture)
C: 1 9: F1 I5 V2
Open correction
Connect SHORT standard.
Press START to continue.
IET Labs.
1701-01 Test
HIGH
LOW
Leads shorted and clips
positioned as shown
Fixture shorted , shorting
bar inserted
C: 1 9: F1 I5 V2
Open correction
Indicates progression of
the cal routine, counts up
Indicates # of record in
table which controls cal
Indicates index of frequency
(0 for DC, 1-14 for ac)
Index of current range
Index of voltage range