Page 94 of 110
Interface
3.4.5 Formats
IEEE 488.2 enable remote programming of all instrument functions, measurement conditions
and comparator settings etc. Outputs include measurement conditions, open corrections, and
measured values.
Data Formats
Data will be transmitted in ASCII NR3 format per IEEE488.2 sec. 8.7.4 and reproduced below.
Multiple results
All response messages will be terminated by the NL character together with the EOI line
asserted.
Status Byte Register
Decimal
Bit
Value
Use
7
128
Measure
completed
6
64
SRQ,
SPOL
Resets
5
32
Summary
of
Standard
Event
Status
Register*
4
16
Message
Available
3
8
Pass
2
4
None
1
2
None
0
1
None
*The Status Byte Register is readable via the standard *STB? as defined in paragraph. 11.2.2.2
of the IEEE spec. The 1910 will also implement an SRE register to enable each bit of the Status
Byte Register per paragraph 11.3.2 of the IEEE spec. This register shall be readable by a SRE?
command and writeable by a SRE <#> command.
Summary of Contents for 1910
Page 8: ...Page 4 of 110...
Page 10: ...Page 6 of 110...
Page 34: ......
Page 100: ......
Page 107: ...Theory Page 103 of 110 Figure 4 4 Detector Circuits Es IL Ex PL PH Voltage Current...
Page 108: ...Page 104 of 110 Theory Figure 4 5 Digital Signal Processor Es Ex CPU DSP A D Vin A Vin B...