PRS-300 Series
PRS-DMM SOFTWARE
28
Chapter 9
PRS-DMM Self Adjusting Software
9.1 PRS-DMM Self Adjusting
Software
9.1.1 Overview
The PRS-DMM Self Adjusting Software is
a Windows based software package using Lab-
VIEW 2013 RTE.
The software is designed to be used with
the PRS-300 Programmable Resistance Substi-
tuter and the Keysight 3458A or Fluke 8508A
DMMs.
The DMM is used to measure the resistance
of the PRS-300, and then the software adjusts
the output resistance of the PRS-300 to achieve
the closest possible resistance value to the target
resistance value.
This can improve the accuracy of the PRS-
300 from ±(70 ppm + 1 m
Ω
) to better than ±10
ppm short-term without switching.
9.1.2 General Requirements
The PRS-DMM Self Adjusting Software
does require:
• Windows 8 or Windows 10 PC
• National Instruments GPIB-USB-HS or
equivalent GPIB interface
• Fluke 8508A DMM or Keysight 3458A
DMM
• PRS-300 Programmable Resistance
Substituter
• Kelvin cables between PRS-300 and
Fluke 8508A
• GPIB cables as necessary
9.1.3 Software Installation
Install the PRS-DMM Self Adjusting
Software using the supplied USB stick. This
will install the PRS-DMM software as well as
the LabVIEW 2013 RTE. This software is also
available at http://www.ietlabs.com/prs-dmm.
html
If using the PRS-300 with standard USB
interface, the FTDI driver should also be in-
stalled. These can be downloard at http://www.
ietlabs.com/Drivers/FTDI.zip
9.1.4 Connection between DMM and
PRS-300
Connect the PRS-300 to the DMM using
good quality low thermal EMF cables using a 4
terminal kelvin connection.
If the end use of the PRS-300 will be 2
Wire then keep shorting links installed. In the
menu of the PRS-300 it should be con
fi
gured for
“2W” Mode.
If the end use of the PRS-300 will be 4 wire
then remove shorting links. In the menu of the
PRS-300 it should be con
fi
gured for “4W” mode.
See section 3.4.10 for more information on
2-wire and 4-wire connections.
9.1.5 Connections of PRS and DMM
Connect the 3458A or 8508A DMM to the
Summary of Contents for PRS-300 Series
Page 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Page 16: ...SPECIFICATIONS PRS 300 Series 6...
Page 24: ...PRS 300 Series IEEE INTERFACE OPTION 14...
Page 34: ...PROGRAMMING PRS 300 Series 24...
Page 42: ...PRS 300 Series PRS DMM SOFTWARE 32...
Page 50: ...PRS 300 Series APPENDIX B 40...