background image

3

SRL Series

3

Specifi cations

 Nominal 

Value 

 Model 

Number 

 Adjustment to 

Nominal 

Stability per 

year (max 

change)

 Max Resistance 

Change 18-28°C 

from 23 °C 

Max Applied Power

Typical change 

at 1 kHz

 Terminals 

 0 ppm change* 

 1 ppm change** 

 3 ppm change** 

 1 mΩ 

 SRL-0.001 

 ±100 ppm 

 ±50 ppm 

 25 ppm/°C

<100 ppm

 4 bp's + gnd 

 10 mΩ 

 SRL-0.01 

 ±5 ppm 

 ±15 ppm 

 5 ppm/°C 

 25 mW 

 50 mW 

 200 mW 

 4 bp's + gnd 

 19 mΩ 

 SRL-0.019 

 ±5 ppm 

 ±15 ppm 

 5 ppm/°C 

 25 mW 

 50 mW 

 200 mW 

 4 bp's + gnd 

 20 mΩ 

 SRL-0.02 

 ±5 ppm 

 ±15 ppm 

 5 ppm/°C 

 25 mW 

 50 mW 

 200 mW 

 4 bp's + gnd 

 100 mΩ 

 SRL-0.1 

 ±5 ppm 

 ±12 ppm 

 2 ppm/°C 

 50 mW 

 100 mW 

 250 mW 

 4 bp's + gnd 

 190 mΩ 

 SRL-0.19 

 ±5 ppm 

 ±12 ppm 

 2 ppm/°C 

 50 mW 

 100 mW 

 250 mW 

 4 bp's + gnd 

 200 mΩ 

 SRL-0.2 

 ±5 ppm 

 ±12 ppm 

 2 ppm/°C 

 50 mW 

 100 mW 

 250 mW 

 4 bp's + gnd 

 1 Ω 

 SRL-1 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 175 mW 

 350 mW 

 850 mW 

 4 bp's + gnd 

 1.9 Ω 

 SRL-1.9 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 175 mW 

 350 mW 

 850 mW 

 4 bp's + gnd 

 2 Ω 

 SRL-2 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 175 mW 

 350 mW 

 850 mW 

 4 bp's + gnd 

 10 Ω 

 SRL-10 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 19 Ω 

 SRL-19 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 20 Ω 

 SRL-20 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 25 Ω 

 SRL-25 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 30 Ω 

 SRL-30 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 50 Ω 

 SRL-50 

 ±2 ppm 

 ±8 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 100 Ω 

 SRL-100 

 ±2 ppm 

 ±6 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 190 Ω 

 SRL-190 

 ±2 ppm 

 ±6 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 200 Ω 

 SRL-200 

 ±2 ppm 

 ±6 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 350 Ω 

 SRL-350 

 ±2 ppm 

 ±6 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 400 Ω 

 SRL-400 

 ±2 ppm 

 ±6 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 1 kΩ 

 SRL-1K 

 ±2 ppm 

 ±6 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 1 kΩ 

 SRL-1K -TC

 ±2 ppm 

 ±6 ppm 

2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 1.9 kΩ 

 SRL-1.9K 

 ±2 ppm 

 ±6 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 2 kΩ 

 SRL-2K 

 ±2 ppm 

 ±6 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 4 kΩ 

 SRL-4K 

 ±2 ppm 

 ±4 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 10 kΩ 

 SRL-10K 

 ±2 ppm 

 ±4 ppm 

 1.5 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 19 kΩ 

 SRL-19K 

 ±2 ppm 

 ±4 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 20 kΩ 

 SRL-20K 

 ±2 ppm 

 ±4 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 100 kΩ 

 SRL-100K 

 ±2 ppm 

 ±6 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

<200 ppm

 4 bp's + gnd 

 190 kΩ 

 SRL-190K 

 ±2 ppm 

 ±8 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 200 kΩ 

 SRL-200K 

 ±2 ppm 

 ±8 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 4 bp's + gnd 

 1 MΩ 

 SRL-1M 

 ±2 ppm 

 ±8 ppm 

 2 ppm tot 

 100 mW 

 200 mW 

 500 mW 

<1000 ppm

 2 bp's + gnd 

 1.9 MΩ 

 SRL-1.9M 

 ±2 ppm 

 ±9 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 2 bp's + gnd 

 2 MΩ 

 SRL-2M 

 ±2 ppm 

 ±9 ppm 

 3 ppm tot 

 100 mW 

 200 mW 

 500 mW 

 2 bp's + gnd 

 10 MΩ 

 SRL-10M 

 ±2 ppm 

 ±9 ppm 

 3 ppm tot 

 500 V 

 1000 V 

 2500 V 

<2 %

 2 bp's + gnd 

 19 MΩ 

 SRL-19M 

 ±2 ppm 

 ±10 ppm 

 4 ppm tot 

 1000 V 

 2000 V 

 5000 V 

NA

 2 bp's + gnd 

 20 MΩ 

 SRL-20M 

 ±2 ppm 

 ±10 ppm 

 4 ppm tot 

 1000 V 

 2000 V 

 5000 V 

 2 bp's + gnd 

 100 MΩ 

 SRL-100M 

 ±10 ppm 

 ±20 ppm 

 5 ppm/°C 

 2000 V 

 4000 V 

 5000 V 

 2 bp's + gnd + guard 

 190 MΩ 

 SRL-190M 

 ±10 ppm 

 ±20 ppm 

 5 ppm/°C 

 2000 V 

 5000 V 

  

 2 bp's + gnd + guard 

 200 MΩ 

 SRL-200M 

 ±10 ppm 

 ±20 ppm 

 5 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 1 GΩ 

 SRL-1G 

 ±0.1% 

 ±200 ppm 

 23 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 1.9 GΩ 

 SRL-1.9G 

 ±0.1% 

 ±200 ppm 

 23 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 2 GΩ 

 SRL-2G 

 ±0.1% 

 ±200 ppm 

 23 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 10 GΩ 

 SRL-10G 

 ±0.1% 

 ±500 ppm 

 25 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 19 GΩ 

 SRL-19G 

 ±0.1% 

 ±500 ppm 

 25 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 20 GΩ 

 SRL-20G 

 ±0.1% 

 ±500 ppm 

 25 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 100 GΩ 

 SRL-100G 

 ±0.3% 

 ±500 ppm 

 25 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 190 GΩ 

 SRL-190G 

 ±0.3% 

 ±500 ppm 

 25 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 200 GΩ 

 SRL-200G 

 ±0.3% 

 ±500 ppm 

 25 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 1 TΩ 

 SRL-1T 

 ±0.5% 

 ±500 ppm 

 50 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 1.9 TΩ 

 SRL-1.9T 

 ±0.7% 

 ±1000 ppm 

 100 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 2 TΩ 

 SRL-2T 

 ±0.7% 

 ±1000 ppm 

 100 ppm/°C 

 5000 V 

  

  

 2 bp's + gnd + guard 

 XXX Ω 

 SRL-XXX 

 customer-selected value and power specifi cations 

 * negligible effect of self-heating; do not exceed voltage limits where given. 

 ** non-permanent self-heating change; exceeding power given for 3 ppm change may cause a permanent change in the resistance. 

Table 2-1:  SRL Specifi cations

Summary of Contents for SRL Series

Page 1: ...ASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC SRL Series Resistance Standard Operation Manual Copyright 2022 IET Labs Inc Visit www ietlabs com for ma...

Page 2: ...PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Page 3: ...t no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci cations shall void this warranty IET shall n...

Page 4: ...ctive 2002 96 EC This product complies with the WEEE Directive 2002 96 EC marking requirements The a xed label indicates that you must not discard this electrical electronic product in domestic househ...

Page 5: ...ions are taken Make all connections to the instrument before applying power Note the instrument s external markings described under Safety Symbols DO NOT Operate in an Explosive Atmosphere Do not oper...

Page 6: ...USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working w...

Page 7: ...ons for values 100 M 5 3 3 Thermal emf Considerations 5 3 4 Temperature Coe cient Constants 6 3 5 Environmental Conditions 6 3 5 1 Operating Temperature 6 3 5 2 Storage Temperature 6 3 6 Shipping and...

Page 8: ...lues 190 kW 4 Table 3 1 Connections for values 190 kW 4 Figure 3 2 Connections for values 1 MW and 100 MW 4 Table 3 2 Connections for values 1 MW and 100 MW 4 Figure 3 3 Connections for values 100 MW...

Page 9: ...s the accuracy by indicating the deviation from nominal for the operating temperature range in 0 5 C increments Because of the low temperature coe cient they require no oil or temperature bath The SRL...

Page 10: ...5 cm W x 12 7 cm D 3 4 x 4 15 x 5 Weight 0 73 kg 1 6 lb Figure 2 1 OPERATION GUIDE affixed to unit Figure 2 2 Temperature Calibration Chart JOS 19 Jun 2002 SRL 1 B2 9240246 N A 1 000 003 3 N A N A SRL...

Page 11: ...ppm 4 ppm 2 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 10 k SRL 10K 2 ppm 4 ppm 1 5 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 19 k SRL 19K 2 ppm 4 ppm 2 ppm tot 100 mW 200 mW 500 mW 4 bp s gnd 20 k SRL 20...

Page 12: ...esistance Change 18 C to 28 C 3 ppm Maximum Voltage 0 5 V Maximum Current 0 5 A LO HI LO HI SENSE CURRENT GND HI INPUT HI SENSE LO INPUT LO SENSE GROUND Binding Post Function CURRENT HI Current input...

Page 13: ...HI GROUND GUARD HI INPUT HI SENSE LO SENSE LO INPUT GROUND GUARD If no GUARD point exists on the measuring instrument it may be connected to GROUND 3 3 Thermal emf Considerations High quality gold pl...

Page 14: ...ult the Temperature Calibration Chart provided with each unit The calculated resistance value is provided be tween 18 C and 28 C in 0 5 C increments Figure 2 2 shows an example of this table 3 5 2 Sto...

Page 15: ...e device the following should be observed Calibration Interval General Considerations Required Equipment Calibration Procedure 4 3 1 Calibration Interval The recommended SRL Series calibration interva...

Page 16: ...precision high stability digi tal multimeter e g Fluke 8508A along with a set of resistance stan dards for ratio mode 4 3 4 Calibration Procedure To calibrate an SRL unit proceed as follows 1 Set up t...

Page 17: ......