5.2 Installation conditions
To achieve the specified measurement accuracy, the following mounting condi-
tions must be adhered to: defined inflow / outflow pipe lengths, defined flow cross-
sectional area, fixed installation depth and correct positioning of the measuring
elements�
The unit is delivered mounted on a pipe section meeting these conditions�
In case of turbulances at the inflow side additional baffled pipes (B) are recom-
mended:
changes to the pipe diameter
B = 5 x pipe diameter
90° elbow
B = 5 x pipe diameter
two 90° elbows, one plane
B = 10 x pipe diameter
two 90° elbows, two planes
B = 15 x pipe diameter
valve, slide
B = 35 x pipe diameter