6
The IO-Link interface provides the following functions:
• Remote parameter setting of the unit�
• Noise immune process value transmission without measured value losses�
• Transmission of the parameter settings to other units of the same type or when
the unit is replaced�
• Simultaneous reading of the two temperature values (measuring element and
reference element), the diagnostic output (calibration check function) and the
device status�
• Paperless logging of parameter sets, process values and diagnostic
information�
5.2 Diagnostic signal
The diagnostic output is only used to provide warnings and alarms (→ data sheet
for the maximum nominal current of the switching output)�
The switching output is closed in normal operation (normally closed)� If the unit
detects a diagnostic case, the output opens�
The switching logic (PnP/nPn) and the calibration check limit (ccL) from which the
output switches can be set�
The diagnostic messages are defined and cannot be changed
5.3 Analogue signal
The unit converts the measured temperature value into a temperature-proportional
analogue signal�
In addition the analogue output is used for diagnostics
�
The transmission of the measured temperature value is interrupted for alarms
and an analogue signal according to NE43 of 3�5 mA (FOU2 = OFF) or 21�5 mA
(FOU2 = On) is provided� In case of warnings, the analogue signal continues to be
provided�
Depending on the parameter setting
the analogue signal for values within
the measuring range is:
4���20 mA with setting [OU2] = I or
20���4 mA with setting [OU2] = Ineg�