Rev. 1.0
Page 36
START
STOP
Measuring Filters
Interconnect:
1. Connect an RF cable to the COM-120B’s AUX RF Output.
2. Connect an RF cable to the COM-120B’s Antenna Input.
3. Using a BNC Barrel, connect the two RF cables together forming a
loop with two cables from the AUX RF Output to the Antenna Input of
the COM-120B. This will configure the COM-120B to zero out test
cables for making insertion loss measurements.
Setup:
1. This COM-120B has been pre-configured for common Analyzer test
setups.
2. Press the
key to access the COM-120B System Setup
Utility software.
3. The menu to the right will appear.
4. Select
STEP 8
by positioning the
cursor and pressing
F1
Execute
.
5.
The COM-120B will configure itself
for filter measurements.
1
2
Summary of Contents for COM-120B
Page 1: ...Rev 1 0 Page 1...
Page 2: ...Rev 1 0 Page 2...