I N T E R N A L C H A N N E L S S E T T I N G S
24
MRBP.424321.001G2 Version 1.0 15/06/2020
5.2.7
«
THRESHOLD 1 HYSTERESIS
»/ «
THRESHOLD 2 HYSTERESIS
»
Hysteresis value is set in the range from 0 to the value not exceeding threshold value.
Edited category is chosen by using the buttons
and
.
The category is changed by using the buttons
and
.
5.2.8
«
THRESHOLD 1 RESET
»/ «
THRESHOLD 2 RESET
»
In the
«
THRESHOLD RESET
»
menu it is possible to select automatic or manual threshold re-
set type. Manual threshold reset is carried out by using handshaking button
. Automatic reset is
carried out after current concentration value decreasing below threshold level.
Notice:
If threshold is exceeded on more than one channel, pressing handshaking but-
ton actuates reset of all the channels with manual threshold reset.
5.3
CHANNEL INPUT CALIBRATION
To calibrate channel input, it is necessary to enter
«
MENU
»
→ «
CHANNELS SETTINGS
»
→
«
INTERNAL CHANNELS
» →
«
CHAN.0
Х»
→
«
INPUT CALIBRATION
».
CH1: INPUT CALIBRATION
CODE 4 mA: < 04896 >
CODE 20 mA: < 24480 >
SAVE
CURRENT ADC CODE: 4896
CURRENT, mA: 4,000
In the
«
INPUT CALIBRATION
»
menu it is possible to calibrate current loop measurement