If the determined blood pressure (systolic or diastolic) is outside the rated range
specified in the SPECIFICATIONS, the App will immediately display a technical alarm.
The technical alarm is preset in the factory and cannot be adjusted or inactivated.
This technical alarm is assigned as low priority according to IEC 60601-1-8. The tech-
nical alarm is non-latching and does not need to be reset.
When the monitor is connected to an iOS device, the battery volume will be dis-
played on the iOS device. It is suggested that you charge the battery when the battery
is less than 25%. The monitor will not work until the battery has enough power.
A medical AC adapter with an output of DC 5.0V and compliant with IEC 60601-
1/UL 60601-1 and IEC 60601-1-2/EN 60601-1-2 is suitable for this monitor, such as
ASP5-05010002EU (input: 100-240V, 50/60Hz, 200mA; output: DC 5V, 1.0A). Please
note that the monitor jack size is USB mini B.
This product might not meet its performance specifications if stored or used out-
side the specified temperature and humidity ranges.
Please do not use the cuff with any infectious person to avoid cross-infection.
This device is designed for adults and should not be used on infants, young chil-
dren, pregnant or pre-eclamptic patients.
Please do not use any other cuff than that supplied by the manufacturer as this
may result in measurement errors and a biocompatible hazard.
BATTERY HANDLING AND USAGE
•
When charging is needed, please connect the monitor to a power source.
•
Do not change the battery. If the battery can no longer be charged, please contact
the iHealth Customer Service.
•
Overcharging the battery may reduce its lifetime.
- The coefficient of variation (CV) of pulse period >25%.
- The difference of adjacent pulse period ≥0.14s and the number of such pulse takes
more than 53 percent of the total number of pulses.
For information regarding potential electromagnetic or other interference be-
tween the blood pressure monitor and other devices, together with advice regarding
avoidance of such interference, please see ELECTROMAGNETIC COMPATIBILITY IN-
FORMATION. It is suggested that the device be kept 10 meters away from other wire-
less devices, such as WLAN unit, microwave oven, etc.
This product is verified according to the auscultatory method. Please check Annex
B of ANSI/AAMI SP-10:2002+A1:2003+A2:2006 for verification method details if
needed.