are designated for repair can be supplied.
•
Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth
using water,diluted disinfectant alcohol, or diluted detergent.
•
The monitor can maintain the safety and performance characteristics for a mini-
mum of 10,000 measurements or three y
performance characteristics for a minimum of 1000 measurements.
•
The battery can maintain the performance characteristics for a minimum of 300
charge cycles.
•
•
Do not drop this device or subject it to strong impact.
•
Avoid high temperature and direct sunlight. Do not immerse the device in water
as this will result in damage to the monitor.
•
Do not attempt to disassemble this device.
•
Battery replacement should only be performed by a qualified iHealth technician.
•
C
•
EXPLANATION OF SYMBOLS
Symbol for “TYPE BF APPLIED PARTS” (C
Symbol for ”THE OPERATION GUIDE MUST BE READ”
(The sign background color:blue The sign graphical symbol:white)
CARE AND MAINTENANCE
•
If this monitor is stored near freezing temperatures, allow it to acclimate to room
temperature before use.
•
If the monitor is not used for a long time, please fully charge it before use.
•
It is recommended that product performance be checked every 2 years or after
each repair. Please contact the iHealth Customer Service for this.
•
No monitor component needs to be maintained by the user. The circuit diagrams,
component part lists, descriptions, calibration instructions, or other information which
will assist qualified technical personnel to repair those parts of the equipment that
13