10
The monitor, cable, battery and cuff must be disposed of
according to local regulations at the end of their usage.
Note
:
The battery has limited charge cycles and may eventually
need to replaced by an iHealth service provider. Battery life
and charge cycles vary by use and settings.
Monitor
Status
Status Indicator
Charging
dynamical battery symbol
Fully
charged
full battery symbol
Low battery empty battery symbol and reminder”Low battery
unable to perform measurement, please
recharge”
TECHNICAL ALARM DESCRIPTION
The monitor will show 'SYS or DIA beyond measurement range' or
' SYS or DIA below measurement range'' as technical alarm on screen
with no delay if the determined blood pressure (systolic or
diastolic) is outside the rated range specified in part
SPECIFICACIONS. In this case, you should consult a physician or
check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset
in the factory and cannot be adjusted or inactivated. This alarm
condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal
displayed on screen will disappear automatically after about 8
seconds.
TROUBLESHOOTING
PROBLEM POSSIBLE
CAUSE
SOLUTION
Low Battery
Battery is less
than 20%
Charge the battery
Display reads
“ERR xxx”
Blood pressure is
outside of
measurement range
Retest, make sure your
blood pressure is within
measurement range
Arm or monitor was
moved during test
Retest, make sure not to
move your arm or the
monitor
The cuff does not
inflate properly
or pressure falls
quickly during
test
Review the cuff
applizcation instructions
and retest