background image

 

7

15.

 

Environmental temperature for storage and transport: -20

Ԩ

50

Ԩ

(-4

Ԭ

122

Ԭ

16.

 

Environmental humidity for storage and transport: 

85%RH 

17.

 

Environmental pressure: 80kPa-105kPa 

18.

 

Battery life: more than 180 measurements on a full charge 

19.

 

The blood pressure measurement system includes accessories, pump, 

valve, cuff, and sensor. 
 

Note: These specifications are subject to change without notice. 

 
 

GENERAL SAFETY AND PRECAUTIONS 

 

1.

 

Read all of the information in the Owner’s Manual and other provided 
instructions before operating the unit. 

2.

 

Stay still, calm and rest for 5 minutes before blood pressure 
measurement. 

3.

 

The cuff should be placed at the same level as your heart. 

4.

 

During measurement, neither speak nor move your body and arm. 

5.

 

Measuring on same arm for each measurement. 

6.

 

Please always relax at least 1 or 1.5 minutes between measurements 
to allow the blood circulation in your arm to recover. Prolonged 
over-inflation (cuff pressure exceed 300 mmHg or maintained above15 
mmHg for longer than 3 minutes) of the bladder may cause ecchymoma 
of your arm. 

7.

 

Consult your physician for any of the following situations: 

a) The application of the cuff over a wound or inflamed area 

b) The application of the cuff on any limb with intravascular access 
or therapy, or an arterio-venous (A-V) shunt  

c) The application of the cuff on the arm on the side of a mastectomy 
d) Simultaneous use with other medical monitoring equipment on the 

same limb 

e) The blood circulation of the user needs to be checked 
 

8.

 

This Blood Pressure Monitor is designed for adults and should never 

be used on infants, young children, pregnant or pre-eclamptic 
patients. Consult your physician before use on children. 

9.

 

Do not use this product in a moving vehicle as this may result in 
inaccurate measurements. 

10.

 

Blood pressure measurements determined by this product are 

equivalent to those obtained by professional healthcare 
practitioners using the cuff/stethoscope auscultation method within 
the limits prescribed by the American National Standard, Electronic 
or Automated Sphygmomanometer. 

Summary of Contents for BPM1AE

Page 1: ...ealth App 3 Set Wi Fi 3 OPERATING PROCEDURES 4 Connecting The Cuff To The Monitor 4 Apply The Cuff 4 Body Posture 5 OPERATION INSTRUCTIONS 5 SPECIFICATIONS 6 GENERAL SAFETY AND PRECAUTIONS 7 BATTERY H...

Page 2: ...ent system It is designed to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the upper arm T...

Page 3: ...th App from the App Store iOS device or Google Play Android device Use keyword search terms iHealth BPM1AE or Blood Pressure Monitor Set Wi Fi Step 1 Press and hold the user2 key for 10s the monitor w...

Page 4: ...f tubing connector the monitor will turn on Apply The Cuff a Pull the cuff end through the metal loop positioning it outward away from your body b Place a bare arm through the cuff and position the cu...

Page 5: ...placed at the same level as your heart Note Blood pressure can be affected by the position of the cuff and your physiologic condition OPERATION INSTRUCTIONS a After applying the cuff and your body is...

Page 6: ...asurements SPECIFICATIONS 1 Product name Blood Pressure Monitor 2 Model BPM1AE 3 Classification Internally powered Type BF applied part IP20 No AP or APG Continuous operation 4 Machine size approx 4 7...

Page 7: ...ressure exceed 300 mmHg or maintained above15 mmHg for longer than 3 minutes of the bladder may cause ecchymoma of your arm 7 Consult your physician for any of the following situations a The applicati...

Page 8: ...ranges 15 Please do not share the cuff with any infectious person to avoid cross infection 16 This product should not be used as a USB device 17 If the determined blood pressure systolic or diastolic...

Page 9: ...charging If the cuff doesn t connect with the monitor press any key the door temperature outdoor tempetature if Wi Fi set done the battery valume and clock if Wi Fi set done will display on monitor If...

Page 10: ...ated range specified in part SPECIFICACIONS In this case you should consult a physician or check if your operation violated the instructions The technical alarm condition outside the rated range is pr...

Page 11: ...f supplication instructions and retest Body posture was not correct during testing Review body posture instructions and retest Speaking moving arm or body being angry excited or nervous during test Re...

Page 12: ...n the safety and performance characteristics for a minimum of 10 000 measurements or three years of usage and the cuff integrity is maintained after 1 000 open close cycles of the closure 10 The batte...

Page 13: ...cal authority or retailer for recycling advice Symbol for MANUFACTURER Symbol for COMPILES WITH MDD93 42 EEC REQUIREMENTS Symbol for DATE OF MANUFACTURE Symbol for SERIAL NUMBER Symbol for EUROPEAN RE...

Page 14: ...oduct does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or m...

Page 15: ...r glementaires et certifications iHealth peuvent tre consult s via ce lien https www ihealthlabs eu support certifications OTHER STANDARDS AND COMPLIANCES The Blood Pressure Monitor corresponds to th...

Page 16: ...munity test levels Home Healthcare Environment Electrostatic Discharge IEC 61000 4 2 8 kV contact 2kV 4kV 8kV 15kV air Radiated RF EM field IEC 61000 4 3 10V m 80MHz 2 7GHz 80 AM at 1kHz Proximity fie...

Page 17: ...kV 100kHz repetition frequency Surges Line to line IEC 61000 4 5 0 5 kV 1 kV Surges Line to ground IEC 61000 4 5 0 5 kV 1 kV 2 kV Conducted disturbances induced by RF fields IEC 61000 4 6 3V 0 15MHz...

Reviews: