8
11.
Information regarding potential electromagnetic or other
interference between the blood pressure monitor and other devices
together with advice regarding avoidance of such interference,
please see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is
suggested that the blood pressure monitor be kept 10 meters away from
other wireless devices, such as WLAN unit, cell phone, microwave oven,
etc.
12.
If Irregular Heartbeat (IHB) is detected during the measurement
procedure, the IHB symbol will be displayed. Under this condition,
the Blood Pressure Monitor can keep functioning, but the results may
be inaccurate. Please consult your physician for accurate
assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period
≥
0.14s and the number
of such pulse takes more than 53 percent of the total number of pulses.
13.
Please do not use any other cuff other than that supplied by the
manufacturer as this may result in measurement errors and a
biocompatible hazard.
14.
This product might not meet its performance specifications if
stored or used outside the specified temperature and humidity ranges.
15.
Please do not share the cuff with any infectious person to avoid
cross-infection.
16.
This product should not be used as a USB device.
17.
If the determined blood pressure (systolic or diastolic) is
outside the rated range specified in part SPECIFICATIONS, the app
will immediately display a technical alarm on screen. In this case,
consult a physician or ensure that proper measurement procedures are
followed. The technical alarm is preset in the factory and cannot
be adjusted or inactivated. This technical alarm is assigned as low
priority according to IEC 60601-1-8.The technical alarm is
non-latching and does not need to be reset.
18.
A medical AC adapter with an output of DC 5.0V and complies with
IEC 60601-1/UL 60601-1 and IEC 60601-1-2/EN 60601-1-2 is suitable
for this monitor. Please note that the monitor jack size is USB
micro-B.
19.
Use of Charging Cable other than those specified or provided by
the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
20.
Measurements are not possible in patients with a high frequency
of arrhythmias.
21.
The device is not intended for use on neonates, children or
pregnant women. (Clinical testing has not been conducted on neonates,
children or pregnant women.)
22.
Motion, trembling, shivering may affect the measurement reading.