LED display reads

“Er4”

Pneumatic system leakage or cuff is too loose during inflation

LED display reads

“Er5”

Cuff pressure above 300mmHg

Measure again after five minutes. If the monitor is still abnormal, please

contact the local distributor or the factory.

LED display reads

“Er6”

More than 160 seconds with cuff pressure above 15 mmHg

LED display reads

“Er7”

memory accessing error

LED display reads

“Er8”

Device parameter checking error

LED display reads

“ErA”

Pressure sensor parameter error

LED display reads

Er”

Bluetooth communicate error

Connect the mobile device correctly and try again, If the monitor is still

abnormal, please contact the local distributor or the factory.

LED display reads

an abnormal

result

The cuff position was not correct or it was not

properly tightened

Review the cuff application instructions and retest

Body posture was not correct during testing

Review body posture instructions and retest

Speaking, moving arm or body, being angry,

excited or nervous during test

Retest when calm; avoid speaking or movement during the test

Bluetooth

connection

unstable

Bluetooth connection unsuccessful, monitor is

abnormal, or strong electromagnetic interference

is present

Reset iOS/Android device. Reset monitor by pressing the START/STOP

button and holding for about 10 seconds. Make sure the monitor and

iOS/Android device are away from other electrical equipment. Please see

GENERAL SAFETY AND PRECAUTIONS

No response when you

press button

Incorrect operation or strong electromagnetic interference

Press the START/STOP button and hold for about 10 seconds to reset the

device.

CARE AND MAINTENANCE

1. If this monitor is stored near freezing temperatures, allow it to return to room temperature before use.

2. If the monitor is not used for a long time, please be sure to fully charge it every month.

3. No monitor component needs to be maintained by the user. The circuit diagrams, component part lists, descriptions, calibration

instructions, or other information which will assist the user’s appropriately qualified technical personnel to repair those parts of

the equipment which are designated for repair can be supplied by the iHealth technical department.

4. Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth using water, diluted disinfectant alcohol, or

diluted detergent.

5. The monitor can maintain the safety and performance characteristics for a minimum of 10,000 measurements or three years of

usage, and the cuff integrity is maintained after 1,000 open close cycles.

6. The battery can maintain the performance characteristics for a minimum of 300 charge cycles.

7. It is recommended that if the cuff is used in a hospital or a clinic, it be disinfected twice a week. Wipe the inner side (the side

that contacts skin) of the cuff with a soft cloth lightly moistened with Ethyl alcohol (75-90%). Then air dry the cuff.

8. It is recommended that product performance be checked every 2 years or after each repair. Please contact the

iHealth Customer Service Center to do so.

Do not drop this monitor or subject it to strong impact.

Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result in damage to the

monitor.

Summary of Contents for View

Page 1: ...T UP REQUIREMENTS 2 SET UP PROCEDURES 3 MEASUREMENT PROCEDURES 3 SPECIFICATIONS 4 GENERAL SAFETY AND PRECAUTIONS 5 BATTERY HANDLING AND USAGE 6 TROUBLESHOOTING 6 CARE AND MAINTENANCE 7 WARRANTY INFORM...

Page 2: ...gned to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist The measurement range of t...

Page 3: ...sure measurements 1 Sit comfortably with your feet flat on the floor without crossing your legs Stay still during measurement Do not move your wrist body or the monitor 2 Place your hand palm side up...

Page 4: ...ar and then select it to pair This may take up to 30 seconds and your monitor s Bluetooth indicator will light up when a connection has been established Launch the iHealth MyVitals app to sync with yo...

Page 5: ...ethod within the limits prescribed by the American National Standard Electronic or Automated Sphygmomanometer This device is also clinically validated according to the 2010 Protocol of the European So...

Page 6: ...ery when the battery is less than 25 charged Overcharging the battery may reduce its lifetime When in charging mode the charging status will be displayed on the LED screen See the table below for deta...

Page 7: ...gnetic interference Press the START STOP button and hold for about 10 seconds to reset the device CARE AND MAINTENANCE 1 If this monitor is stored near freezing temperatures allow it to return to room...

Page 8: ...lue The sign graphical symbol white Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your loc...

Page 9: ...easures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receive...

Page 10: ...c discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be a...

Page 11: ...all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellul...

Page 12: ...rs m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufactu...

Reviews: