TECHNICAL SPECIFICATIONS

130

CIC Response

2cc Coupler Response

120

110
100

90
80

100

1000

Frequency [Hz]

dBSPL

10000

CIC Response

2cc Coupler Response

50
40

30

20

10

0

100

Frequency [Hz]

dB

1000

10000

MAXIMUM OUTPUT (OSPL 90)

ANSI S3.22 - 2003

FREQUENCY RESPONSE & MAXIMUM GAIN*

AT 60 dBSPL INPUT

*Nominal compression ratio settings of 1.4

O

utput

Maximum Output

HF Average OSPL90

A

mplificAtiOn

Max. Peak Gain

HF Average RTG

Minimum Frequency Range

H

ArmOnic

 D

istOrtiOn

500 Hz
800 Hz

1600 Hz

E

quivAlEnt

 i

nput

 n

OisE

A

DAptivE

 A

ttAck

 t

imE

A

DAptivE

 r

ElEAsE

 t

imE

109 dB
104 dB

27 dB
23 dB
200 Hz - 7 kHz

1.5%
1.5%
1.5%
20 dBSPL
8 ms - 256 ms
32 ms - 256 ms

Computer and Software Requirements*

u

  

Personal computer with USB 2.0+ port

u

  

iHear Software (Available for download online)

u

  

Operating System:   Windows 7 or higher  

 

Mac OS 10.9 or higher

u

  

Internet connection 512 Kbps or higher

*Visit 

www.ihearmedical.com/faq

 for the most  

 up-to-date compatibility information.

Summary of Contents for IHEAR HD

Page 1: ...HIGH DEFINITION HEARING DEVICE User Guide iHear Loud Clear...

Page 2: ...under the age of 18 The Food and Drug Administration has determined that it is in your best health interest to have a medical evaluation by a licensed physician preferably a physician who specializes...

Page 3: ...earUSB module and the Programming Cable are essential tools used to program your iHEARHD device Open the iHearTest and remove the iHearUSB module Locate the Programming Cable inside the Travel Case RE...

Page 4: ...reate a new account fill in the required information then Click on Sign up Click on Sign In to sign in to an existing account Once you ve created your iHear account you will receive notification by em...

Page 5: ...n and product activation Launch iHTUNER If you have just finished installing the software by default the iHTUNER application should launch automatically for you the first time If you need to re launch...

Page 6: ...Current ear infection History of active drainage from the ear within the past 90 days Sudden or rapid progression of hearing loss within the past 90 days in one or both ears Acute or chronic dizziness...

Page 7: ...icates a secure connection PUSH UP Handle PULL UP Insert the Programming Pill to the iHEARHD device Align the Tab to the Hook Rotate Down Lock Repeat this step if you purchased both Left Right hearing...

Page 8: ...rogramming Cable with the Tip into the Ear Canal Sizing the ear canal Attach the selected Tip to the Programming Cable Select the Correct Comfort Tip Tip Size for your Ear s Start with Medium Too tigh...

Page 9: ...N Store the Programming Cable back in its case The Battery Module is designed to seal the device and act as an ear wax filter PULL tape before use to activate the battery After you successfully progra...

Page 10: ...it should be tight but comfortable u Pulling your ear up or back may help insertion u Left and right ear sizes may vary u Repeat this step if you purchased both Left and Right hearing devices PULL OUT...

Page 11: ...Y WHEN YOU SNAP IN THE BATTERY DEVICE START UP PROGRAM 1 PROGRAM 2 PROGRAM 3 PROGRAM 4 Beeps Remote Control Guide Each tap cycles to the next program setting With each tap you will hear beeps that ind...

Page 12: ...iHEARHD into the device socket 1 PUSH IN LEFT RIGHT 2 UP Removing the iHEARHD from the Tool Kit and Turning it ON SNAP IN the Battery Module to Turn ON PULL OUT You should feel a light snap as the ba...

Page 13: ...he Removal Tool optional Place the removal loop over the device handle Gently pull out the iHEARHD device from the ear canal PUSH IN to store in Tool Kit Using the Removal Tool optional PULL OUT Remov...

Page 14: ...evice s marker dot Marker Dots Match Marker Dots Do Not Match X ATTACH the Insertion tool to the iHEARHD Inserting iHEARHD device in the ear using the Insertion Tool optional INSERT the iHEARHD device...

Page 15: ...l Kit has been designed to turn off and hold your device to prolong battery life Clean the iHEARHD using the Cleaning Wipe u Hold the device body as shown and wipe down the entire surface of the devic...

Page 16: ...the device daily and replace the battery module and Comfort Tip periodically to avoid earwax contamination of the device u The iHEARHD is designed to withstand limited water exposure such as surface s...

Page 17: ...90 Amplification Max Peak Gain HF Average RTG Minimum Frequency Range Harmonic Distortion 500 Hz 800 Hz 1600 Hz Equivalent Input Noise Adaptive Attack Time Adaptive Release Time 109 dB 104 dB 27 dB 23...

Page 18: ...period of one year from the date of shipment of the device If the iHEARHD contains a defect in material or workmanship iHear Medical Inc at its option will repair or replace the device at no charge ex...

Page 19: ...accidentally swallowed Made by iHear Medical in California USA U S and international patents issued and pending 2016 iHear Medical Inc All rights reserved iHEAR iHEARHD iHear Loud Clear iHearTest and...

Reviews: