Nebulizer kit is
assembled incorrectly
Assemble the
nebulizer kit correctly.
Nozzle is blocked
Make sure that the
nozzle is free of
blockages.
Nebulizer kit is tilted at
a sharp angle
Make sure that the
nebulizer kit is not
tilted at an angle
of more than 45
degrees.
The air tube is folded,
creased or damaged or
blocked
Make sure that the
air tube does not
contain kinks. Make
sure that air tube is
free of blockages
Air filter is dirty
Replace the air filter
with a new one.
Note: If the suggested remedy does not solve the problem, do not
try to repair the device. Shut off the device for about 1 hour and
then restart the machine. If the problem persists contact your authorized
service provider. Do not disassemble the unit yourself, no user serviceable
parts inside, risk of electric shock.
Before servicing by authorized service personnel. To completely isolate
the device from the power source unplug the plug from the power source.
TECHNICAL SPECIFICATIONS
Power
220 V AC, 50Hz
Power max
100 VA
Compressor air output
2.5 bar, 8 LPM approximate
Medicine Bottle capacity
8 ml / max, 2ml - 7ml recommended
Nebulization rate
0.2 ml / min - 0,4 ml / min typical
(Weight loss Method)