LNU sensor : Installation & programming
Installation Guide
K0J00016-EN-A02
IJINUS – 25 ZA de Kervidanou 3 – 29300 MELLAC – FRANCE
-
14
Finally, you need to launch a new measure to check the relevance of the result so :
- The green line position used for the measure,
- Values of the echo slightly higher than the power indicator,
- and validate by pressing the Apply button
Calibration of the sensor Case N°3 (complex) – Expert mode
The expert mode is to be used only in complex cases and need user experience and knowledge of
ultrasonic measure data processing. Various parameters and options are then available but won’t be in
this document. You need a password to go on expert mode. Please contact your sales representative or
the after sale services, the password as well as some information will be given.
Temperature recording
You can log the temperature thank to the internal T probe used to compensate the level measurement with ultrasonic.
For information, the sound speed in the air at 10°C is 337,6m/s.