LNU sensor : Installation & programming
Installation Guide
K0J00016-EN-A02
IJINUS
–
25 ZA de Kervidanou 3
–
29300 MELLAC
–
FRANCE
-
28
7.
Procedure for inserting the SIM card
- Unscrew the cover and disconnect the antenna
cable.
-
Remove the battery from its housing and
disconnect.
Press several times on the push button until the LED
does not light up. Throw away the desiccant bags
present in the sensor if they are green)
- Connect the new battery (the connector is keyed)
For ATEX devices, only use A0G00011 battery pack
provided by IJINUS
Press the push button
SW1 to restart the sensor
At restart, the indicator light has to flash
Red/Green, and then the light has to flash
every 10 seconds
-
Insert the new desiccant bags
on the side of the
battery, reconnect the antenna to the connector until
you hear a click, then screw the cover back into
place.
- Tell the software you have done a battery change
by connecting with the software Avelour to your
sensor.
To do so launch the software Avelour, connect to the
sensor or logger.
Then go to the general menu,
> Action
> Notify battery changed