Power UP = indicates the power up mode that is programmed into the instrument.
Power UP Mode: indicates the power up mode that can be programmed into the
instrument.
The power up modes available are ON, OFF or LAST. In the ON mode output will
be supplied on power up of the instrument. In the OFF mode output will NOT be
supplied on power up of the instrument. In the LAST mode output will be
supplied according to the last state the instrument was in prior to power off.
Press the Change soft key to toggle the Power UP Mode to ON/OFF/LAST. To
save the parameter, press the Enter soft key. To cancel the editing of the Power
Up feature press the Esc soft key. When the Enter soft key is pressed the
Power Up parameter is accepted and you transition into the next system
parameter.
4.3.7 Editing Loop Cycle (PROGRAM Mode only)
Use the
,
soft keys to navigate to the Loop Cycle parameter (only selectable
for AC output). Pressing the Edit soft key will provide the following screen:
Loop Cycle = indicates the number of loop cycles that will be performed when the
output is active.
Loop Cycle Range: = indicates the selections available for the Loop Cycle Range.
The options available are 0
– 9999, 0=Cont., 1=Off. The 0 – 9999 selection
programs the instrument to repeat the test cycle x number of times. The 0=Cont.
selection indicates that the test loop will repeat in ∞. The 1=Off selection
indicates that the test loop will perform only one cycle. Use the numeric keypad
to enter in the Loop Cycle Range. Press the Enter soft key to accept the
parameter. To cancel the editing of Loop Cycle Range, press the Esc soft key.
To move to the next system parameter for editing, press the Next or Prev soft key.
The Exit soft key is also available to return to the System parameter screen.
When the Enter soft key is pressed the Loop Cycle Mode is accepted and you
transition into the next system parameter.